タイトル: Expansion of Stent and Lumen Diameters on Follow-up after Carotid Artery Stenting in Patients with Carotid Artery Stenosis.
その他のタイトル: 頸動脈狭窄に対する自己拡張型ステント留置後フォローアップ時のステント径と内腔の検討
著者: Fujimoto, Kenta
Hashimoto, Hiroyuki
Nishiguchi, Mitsuhisa
Inui, Toshitaka
Tani, Naoki
Horiuchi, Kaoru
Chatani, Megumi
Matsuoka, Ryuta
Yaegaki, Takahide
Nakagawa, Ichiro
Nakase, Hiroyuki
キーワード: carotid artery stenting
self-expansion
plaque
発行日: 2017年2月
出版者: The Japanese Society for Neuroendovascular Therapy
引用: Journal of Neuroendovascular Therapy Vol.11 No.2 p.65-75 (2017 Feb)
抄録: Purpose: We examined postoperative stent and lumen expansions after carotid artery stenting (CAS) in patients with carotid artery stenosis. Furthermore, we investigated factors influencing the stent and lumen expansions in a follow-up period. Subjects: 134 cases (128 patients) who underwent CAS and performed follow-up cerebral angiography 12 months after CAS were enrolled into this study. The stenosis rate based on the stent and lumen diameters on follow-up angiography as a percentage of that immediately after CAS was evaluated. Results: Both the stent and lumen diameters were significantly dilated 12 months after CAS (p <0.001). There were no significant stent-type-related differences in the stent expansion rate. In the symptomatic stenosis group, this expansion rate was significantly higher than in the asymptomatic stenosis group (p = 0.02). With respect to the presence or absence of a high signal intensity on time of flight (TOF) magnetic resonance (MR) images, the stent expansion rate was significantly higher in the high signal intensity group (p = 0.006). In patients with a plaque/sternocleidomastoid muscle signal intensity ratio of ≥1.50 on plaque images, it was significantly higher than in those with a value of <1.50 (p = 0.006). However, there were no significant differences in the lumen expansion rate among the groups. Conclusion: Both the stent and vascular lumen were dilated 12 months after CAS. Plaque fragility influenced the stent expansion rate; however, there were no significant factor-related differences in the vascular lumen expansion rate.
内容記述: 博士(医学)・乙第1414号・平成30年3月15日
©2017 The Editorial Committee of Journal of Neuroendovascular Therapy. All rights reserved. This is an open access article distributed under the terms of Creative Commons Attribution License(CC BY-NC-ND 4.0) https://creativecommons.org/licenses/by-nc-nd/4.0/.
URI: http://hdl.handle.net/10564/3430
ISSN: 18824072
DOI: https://doi.org/10.5797/jnet.oa.2016-0038
学位授与番号: 24601B1414
学位授与年月日: 2018-03-15
学位名: 博士(医学)
学位授与機関: 奈良県立医科大学
出現コレクション:2017年度

このアイテムのファイル:

ファイル 記述 サイズフォーマット
01乙1414本文の要旨.pdf乙1414本文の要旨1.76 MBAdobe PDF見る/開く
02乙1414審査要旨.pdf乙1414審査要旨324.16 kBAdobe PDF見る/開く
03乙1414本文.pdf乙1414本文1.82 MBAdobe PDF見る/開く

このリポジトリに保管されているアイテムは、他に指定されている場合を除き、著作権により保護されています。

 

Valid XHTML 1.0! Powered by DSpace Software Copyright © 2002-2007 MIT and Hewlett-Packard - ご意見をお寄せください